Performance Benchmark Metrics and Clinicopathologic Outcomes of MRI-Guided Breast Biopsies: A Systematic Review and Meta-Analysis
PDF
Cite
Share
Request
Original Article
P: 1-27
January 2023

Performance Benchmark Metrics and Clinicopathologic Outcomes of MRI-Guided Breast Biopsies: A Systematic Review and Meta-Analysis

Eur J Breast Health 2023;19(1):1-27
1. Department of Radiology, University of Texas Southwestern Medical Center, Dallas, TX, USA
2. Department of Population and Data Sciences, University of Texas Southwestern Medical Center, Dallas, TX, USA
3. Department of Radiology, Cleveland Clinic Foundation, Cleveland, OH, USA
4. Department of Radiology, University of Texas MD Anderson Cancer Center, Houston, TX, USA
No information available.
No information available
Received Date: 05.12.2022
Accepted Date: 22.12.2022
Publish Date: 01.01.2023
PDF
Cite
Share
Request

ABSTRACT

Objective:

To determine key performance metrics of magnetic resonance imaging (MRI)-guided breast biopsies (MRGB) to help identify reference benchmarks.

Materials and Methods:

We identified studies reporting MRGB results up to 04.01.2021 in the Embase database, Ovid Medline (R) Process, Other Non-Indexed Citations, Ovid Medline (R) and completed a PRISMA checklist and sources of bias (QUADAS-2). The inclusion criteria were English language, available histopathological outcomes, or at least one imaging follow-up after biopsy. A random intercept logistic regression model was used to pool rates. Between-study heterogeneity was quantified by the I2 statistic.

Results:

A total of 11,215 lesions in 50 articles were analyzed. The technical success rate was 99.10% [95% confidence interval (CI): 97.89–99.62%]. The MRI indications were staging in 1,496 (28.05%, 95% CI: 26.85–29.28%), screening in 1,427 (26.76%, 95% CI: 25.57–27.97%), surveillance in 1,027 (19.26%, 95% CI: 18.21–20.34%), diagnostic in 1,038 (19.46%, 95% CI: 18.41–20.55%), unknown primary in 74 (1.39%, 95% CI: 1.09–1.74%), and other in 271 (5.08%, 95% CI: 4.51–5.71%). Histopathology was benign in 65.06% (95% CI: 59.15–70.54%), malignant in 29.64% (95% CI: 23.58–36.52%) and high risk in 16.69% (95% CI: 9.96–26.64%). Detection of malignancy was significantly lower in those patients who underwent MRI for screening purposes (odds ratio 0.47, 95% CI: 0.25–0.87; p = 0.02), while mass lesions were more likely to yield malignancy compared to non-mass and foci [27.39% vs 11.36% (non-mass),18.03% (foci); p<0.001]. Surgical upgrade to invasive cancer occurred in 12.24% of ductal carcinoma in situ (95% CI: 7.76–18.77%) and malignancy in 15.14% of high-risk lesions (95% CI: 10.69–21.17%). MRI follow-up was performed in 1,651 (20.92%) patients after benign results [median=25 months (range: 0.4–117)]. Radiology-pathology discordance (2.48%, 95% CI: 1.62–3.77%), false negative after a benign-concordant biopsy (0.75%, 95% CI: 0.34–1.62%) and biopsy complications (2.36%, 95% CI: 2.03–2.72%) were rare.

Conclusion:

MRGB is a highly accurate minimally-invasive diagnostic technique with low false-negative and complication rates. MRI indication and lesion type should be considered when evaluating the performance of institutional MRGB programs.

Keywords:
Breast, cancer, magnetic resonance imaging, biopsy

Key Points

• Magnetic resonance imaging (MRI)-guided breast biopsy methods and clinicopathological outcomes may vary between institutions.

• MRI-guided breast biopsy is an efficient, highly accurate technique with high technical success [99.10%, 95% confidence interval (CI): 97.89–99.62%], low false-negative (0.75%, 95% CI: 0.34–1.62%), and low complication (2.36%, 95% CI: 2.03–2.72%) rates.

• The surgical upgrade to malignancy is common among high-risk lesions 15.14% (95% CI: 10.69-21.17%), especially atypical ductal hyperplasia (31.81% (95% CI: 25.57-38.77%).

Introduction

Magnetic resonance imaging (MRI) has a high sensitivity (88–92%) and a moderate specificity (67–77%) for the detection of breast cancer (1). It has been well established that MRI-guided tissue sampling is necessary for the histological verification of lesions that are otherwise occult (1, 2, 3, 4, 5). Furthermore, due to the overlap of the MRI findings of the benign and malignant lesions, in order to distinguish between them, an MRI-guided breast biopsy is necessary (6).

Surgical biopsy after MRI-guided wire localization and MRI-guided percutaneous needle biopsies have been described before the first experiences with MRI-guided vacuum-assisted biopsy were reported in the late 1990s (7, 8). Since then, MRI-guided vacuum-assisted biopsy has achieved broad acceptance in clinical practice due to its speed, accuracy, and safety, which has been found to be as good as MRI-guided wire localization without the associated complications and cost of surgery (7, 8, 9, 10, 11, 12, 13). MRI-guided needle biopsy also allows for the placement of marker clips and so aids the subsequent mammographic localization of the lesion if an operation becomes necessary.

Tissue sampling with fine-needle aspiration and core-needle biopsy devices requires visual confirmation of needle placement directly into the target to ensure accurate sampling. The suction of the MRI-guided vacuum-assisted biopsy device provides for adequate sampling when the needle is placed within a few millimeters of a small lesion, provided that the suction chamber is preferentially directed toward the target. Thus, the use of vacuum assistance has allowed for the accurate targeting of smaller lesions. In addition, because the vacuum system continuously suctions any hemorrhage which may occur during sampling, tissue shift and subsequent sampling errors are mitigated.

MRI-guided breast biopsy can be a challenging procedure for radiologists. Determining radiologic-pathologic concordance for MRI-guided biopsies is often more difficult than biopsies performed using other imaging modalities. Since it is not a real-time procedure, it lacks the direct needle visualization advantage of ultrasound-guided biopsies. Unlike stereotactic biopsies where intra-procedure specimen radiographs ensure the accuracy of targeting, ex vivo confirmation of sampling is not possible. Furthermore, wash-out of the gadolinium contrast agents during the procedure and post-biopsy changes including air, hemorrhage, and local anesthesia obscure the targeted lesion, making it more challenging to confirm the accuracy of sampling. It is a procedure which obligates sliding the table on the gantry to place the guiding system and performing the biopsy again, without real-time visualization of the lesion. These factors render radiologic-pathologic correlation critical. Lesion enhancement is another challenge while performing MRI-guided biopsy because lesion conspicuity decreases with time after contrast injection due to the enhancement kinetics. Compression of the breast needs to be adequate to immobilize the breast and to ensure hemostasis without obstructing lesion contrast enhancement.

MRI-guided breast biopsy is a time-consuming and complex procedure which requires specific equipment and expertise. Current MRI-guided breast biopsy methods and subsequent clinicopathological outcomes may vary between institutions. Our goal was to identify benchmark metrics to help define a successful breast MRI-guided biopsy program and guide institutional audits. To accomplish our goal, we identified and systematically reviewed studies in order to determine indications, technical success, histopathological outcomes, false-negatives, and upgrade rates of MRI-guided breast biopsies for institutional referencing.

Materials and Methods

The preferred reporting items for systematic reviews and meta-analyses (PRISMA) guideline was used for reporting (14).

Results

Discussion and Conclusion

MRI-guided breast biopsy is an efficient and highly accurate technique with high technical success (99.10%, 95% CI: 97.89–99.62%) and a low false-negative rate (0.75%, 95% CI: 0.34–1.62%). We found a low false-negative rate in benign-concordant lesions, which supports that there is no need to follow-up patients with MRI after a benign-concordant biopsy result (19, 20, 21, 22).

We found that benign biopsies accounted for more than half of all MRI-guided breast biopsies (65.06%, 95% CI: 59.15–70.54%) in all lesion types (60.35% in mass, 61.96% in non-mass enhancement, 68.85% in focus). Our findings suggest that enhancing lesion type by ACR BI-RADS descriptors influenced the malignancy rate and that mass lesions were more likely to yield malignancy compared to non-mass lesions and foci (27.39% vs 11.36% non-mass and 18.03% foci, p<0.001). This finding is in keeping with previous studies which reported that the malignancy rate is higher for masses (34–60%) (23, 24, 25). However, our pooled malignancy rate in mass lesions was somewhat less than had been previously reported. Masses are more likely to be identified on second-look ultrasound (58–65%) than non-mass-like lesions (12–54%), and consequently were more likely to undergo ultrasound-guided needle biopsy (26, 27, 28, 29). It was also reported that lesions which were seen on second-look ultrasound were more likely to be malignant (57.4–91.7%) (28, 29, 30, 31). Collectively, this results in only those masses likely to be benign undergoing MRI-guided biopsy, which can be the reason why we saw a low pooled malignancy rate.

In our study, the pooled rate for malignancy was 29.64% (95% CI: 23.58–36.52%). Patients undergoing MRI for breast cancer surveillance, diagnostic indication, and breast cancer staging had a higher rate of malignant results (OR, 1.36, 1.20, and 1.20; respectively), although none of them were statistically significant (p=0.09–0.40). Detection of malignancy was significantly lower in those patients who underwent MRI for screening purposes (OR 0.47, 95% CI: 0.25, 0.87; p=0.02). When interpreting our results, it should be considered that the study results included were homogeneous in breast cancer surveillance, staging, and diagnostic indication groups whereas in the screening group, they were heterogeneous (p-values of the random effects models were: 0.36, 0.13, 0.71, and 0.03, respectively). In contrast to previous studies which reported the frequency of malignancy to be significantly higher in those patients presenting for diagnostic versus screening purposes (screening 10–14% vs diagnostic 28–36%; p<0.05) (24, 31), we did not compare individual indications with each other. Rather, with a Bayesian model, we compared whether the indication of interest affected the MRI biopsy outcome or not. This difference in analyzing methods should be considered.

ADH identified with MRI-guided biopsy was found to have a pooled underestimation rate of 31.13% (95% CI: 25.17–37.78%), slightly higher than that of stereotactic biopsy (mean 20%, range 10–27%, with 11-gauge vacuum-assisted biopsy probe) (32, 33, 34, 35, 36, 37). ADH has high upgrade rates (15.0–53.3%) verified over multiple studies (31, 38, 39, 40, 41, 42). In a recent study by Michaels et al., it was found that ADH was more likely to upgrade to cancer at surgical excision than other high-risk lesions (22.5% vs 3.4%, p=0.005) and that larger high-risk lesions had a greater tendency for an upgrade than smaller lesions (1.8 vs 1.2 cm, p=0.073). Furthermore, Rauch et al. (31) and Heller et al. (38) reported that the risk of upgrade in MRI-detected high-risk lesions was higher if the high-risk lesion was identified in the same breast as a prior malignancy, or if the patient had had a recent diagnosis of malignancy. Our findings underscore that the surgical upgrade to malignancy is common among high-risk lesions, especially ADH. Traditionally, it has been recommended to surgically remove high-risk lesions due to their high degree of underestimation on biopsy. However, the most recent recommendations advocate a more cautious multidisciplinary approach to assess the individual risk of patients and to avoid surgical excision whenever possible (43, 44). Unfortunately, due to a lack of correlating data on patient history, we could not further investigate multivariable associations on the surgical upgrade of high-risk lesions diagnosed at MRI-guided breast biopsy to predict the individual risks of patients.

MRI-guided breast biopsy is a safe technique with low complication rates (0–6%) (3, 10, 11, 45, 46, 47, 48). Complications are generally minor (hematomas, malaise, skin damage) and easily managed (11, 47, 48). In our systematic review, we found a complication rate of 2.36%, almost all comprising hematomas, and none of them requiring major interventions, such as surgery.

Occasionally, a finding identified as suspicious on prior breast MRI no longer enhances on the day of the biopsy. It has been hypothesized that these cancellations occur as a result of changing hormonal status (related to the menstrual cycle, menopausal status, age, hormone suppression, or replacement therapy) which can affect background parenchymal enhancement, patient positioning, or the over-compression of the breast within the MRI-biopsy coil (11, 24, 49, 50, 51, 52, 53, 54, 55). It has also been reported that non-visualization was more commonly seen in non-mass enhancement (54). In our review, 4.58% (95% CI: 1.81–11.11%) of the scheduled biopsies were canceled due to non-enhancement on the day of the biopsy, with single center reports ranging from 6.9–13% (11, 24, 49, 50, 53, 54, 55). The lower pooled cancellation rate due to non-enhancement in our study may be due to our inclusion of newer studies performed over the last decade, which reflect the learning curve to appropriately recognize normal but variably enhancing parenchyma by radiologists, resulting in fewer biopsies recommended for benign background enhancement. Previously, it had been reported that the cancer detection rate among lesions for which biopsy was canceled due to non-enhancement was low (2–10%) (49, 53, 55). This rate could not be analyzed in our systematic review.

Careful radiologic-pathologic correlation is necessary to confirm the concordance of imaging findings with pathology. In our review, imaging-pathologic discordance occurred in 2.48% (95% CI: 1.62–3.77%) of MRI-guided biopsies. This discordance rate is similar to rates reported for stereotactic and ultrasound-guided needle biopsies (1.3–4.4%) and further validates the MRI-guided breast biopsy technique (56-60). Previously, it was found that lesions which were missed rather than sampled on MRI-guided biopsies had a higher rate of imaging discordance, and lesions with discordant imaging had a higher risk of malignancy (30–100%) (47, 56, 61). This malignancy risk was higher than had been reported for stereotactic-guided biopsy (11.7–53.8%) (58, 59, 60) and ultrasound-guided biopsy (0.1–2.4%) (57, 62, 63). This could have been caused by the MRI patient population characteristics, which includes high-risk patients, patients with newly diagnosed breast cancer or a history of breast cancer. Since a similar discordance rate was observed in MRI-guided biopsy with higher malignancy, there should be a standard reference for reporting false-negative rates in MRI-guided biopsies. We realized that there is no standard of reference and, in some studies, benign-discordant biopsies which were found to be malignant after re-biopsy or surgical excision were counted as false-negatives (64, 65, 66), while in others, those cases were excluded from the false negative cases (67). In our systematic review, we defined the false negative rate as the rate of malignancy identified after a benign-concordant MRI-guided breast biopsy, and the pooled false-negative rate for the studies included was 0.75% (95% CI: 0.34–1.62%).

The limitations of this meta-analysis include the heterogeneity between the groups and the across studies (I-squared >25%). Most studies were retrospective in design, with only three prospective studies contributing data into the pooled estimates. As a result, bias and confounding could not be fully eliminated, and the interpretation of our findings should factor in the heterogeneity between the studies.

In the series published to date, the reported false-negative rates were determined only for those cases in which follow-up or immediate excision/re-biopsy was performed. In addition to that, due to the retrospective study design, only those lesions which were successfully biopsied were reported in some of the studies included (21, 38, 68, 69, 70, 71). Thus, the technical success rate was missing. We did not pool those studies’ data in our technical success rate analysis so as not to inflate the technical success rate. However, the true false-negative and the technical success rates of MR-guided breast biopsy remain to be determined, and this was another limitation of our study.

Most of the articles lacked correlating data between histopathology and clinical indication. Hence, we had to perform our correlation analysis with 9 studies (out of 50 the studies included), which limited the statistical power of our analysis. Another limitation was inconsistent reporting of study-level data for variables such as age, number of cores sampled, lesion sizes, and follow-up times. We used the available mean or median values for those variables in our pooled analysis.

The lack of standardization in reporting the technical success rates and false negative rates made it hard to pool the available data. Despite this, we had determined our reference standards before we began our literature search and stuck to those standards. Three investigators independently extracted the relevant information in addition to reviewing and confirming the selected articles. We also applied Quality Assessment of Diagnostic Accuracy Studies (QUADAS-2) items to assess study quality and the likelihood of bias (15). Our estimates, therefore, represent the most comprehensive evidence summary on breast MRI-guided biopsy outcomes, despite the above-mentioned limitations inherent in this study-level meta-analysis.

MRI-guided breast biopsy is a highly accurate technique with a high technical success rate, and negligible false negative and complication rates. Our findings can be used to guide breast radiologist practice, to inform transparent discussion with patients on the consequences of having an MRI-guided breast biopsy, and to assist the development of evidence-based clinical guidelines on follow-up recommendations in benign-concordant breast lesions. The substantial degree of variation in performance metrics across the studies included in our analysis suggests that ongoing quality improvement efforts are needed.

Literature Search and Article Selection Criteria

The requirement to obtain institutional review board approval was waived for this literature review, which involved only publicly available data. The Ovid MEDLINE® In-Process & Other Non-Indexed Citations, Ovid MEDLINE®, and Embase databases​ were searched systematically for English language articles published from January, 1946 up to April, 2021 for articles on MRI-guided breast biopsy outcomes by an investigator trained in conducting comprehensive literature searches. Three investigators then independently reviewed and confirmed the selected articles and extracted the relevant information.

The search terms included breast neoplasm, MRI/MRI, and image-guided biopsy from articles involving human subjects. The search strategy is shown in Supplementary Table 1.

Table 1

Our inclusion criteria were English language literature, the availability of reported histopathological outcomes of benign, malignant, and high-risk lesions, and the availability of final histopathology (gold standard) or at least one-time imaging/clinical follow-up after biopsy. We excluded meta-analyses, review papers, case-control studies, and matched-pair studies, and included original articles which reported novel data.

We excluded studies that were non-English in their full text, and those where the following information was not reported: Technical factors (magnet strength, needle type, needle gauge), imaging or clinical follow-up descriptions, or time unavailable after a high-risk or benign biopsies. The results of the literature search and applied study selection criteria are summarized in Figure 1.

Figure 1

Data Collection and Quality Assessment

We collected mean/median patient ages, indications for MRI biopsy, magnet strengths, needle types/gauges, the number of cores sampled, rates of successfully performed MRI biopsies, causes of unsuccessful biopsies and pathological outcomes (benign, malignant, high risk) along with false negative rates and underestimation rates for ductal carcinoma in situ (DCIS), atypical ductal hyperplasia (ADH) and other high-risk lesions [lobular carcinoma in situ (LCIS), atypical lobular hyperplasia (ALH), flat epithelial atypical (FEA) radial scar (RSL)/complex sclerosing lesions (CSL)]. The lesion characteristics (mass, non-mass enhancement, focus and size information for each), enhancement kinetics (wash-out, plateau, progressive), complications (if any), and the types and durations of follow-up were also recorded.

One reader applied the modified quality assessment of diagnostic accuracy studies (QUADAS-2) items to assess study quality and the likelihood of bias (15). The risk of bias was judged as “low”, “high” or “unclear” on four domains: Patient selection, index test, reference standard, and flow and timing. Concerns about applicability were judged as “low”, “high” or “unclear” on three domains: Patient population, index test, and reference standard. A study was judged as “at risk of bias” or as having “concerns regarding applicability” when it was judged “high” or “unclear” in one or more domains. A second reader checked the results. If present, disagreement was solved in consensus. Detailed information on signaling questions in each domain is shown in Supplementary Table 2.

Table 2

Our primary outcomes were:

1) Rate of successfully performed MRI biopsies, 2) rate of pathological outcomes of benign, malignant, high-risk, 3) false negative rate, 4) follow-up outcomes after a benign MRI-guided breast biopsy.

We aimed to identify potential technical and patient clinicopathological factors which may have influenced MRI-guided breast biopsy outcomes.

Reference Standards

A false-negative result was defined as a pathologically proven malignancy after follow-up or immediate excision or re-biopsy following an MRI-guided benign biopsy. Discordant biopsy results occur when benign pathology results do not account for the imaging findings and MRI-guided benign histopathology results include both imaging-concordant and -discordant ones. The false-negative rate was defined as the rate of malignancy identified in those patients with benign-concordant MRI-guided breast biopsies.

High-risk lesions were ADH, LCIS, papillary lesions (intraductal papilloma and papilloma with atypia), ALH, FEA, and RSL/CSL (16). High-risk lesions which were diagnosed at MRI-guided biopsy, and in which a subsequent diagnosis of invasive cancer or DCIS lesion was made at surgical excision or follow-up re-biopsy, were considered as underestimations.

The high-risk lesion underestimation rate was defined as the number of these underestimated lesions divided by their high-risk lesion category (ADH vs other high-risk lesions) at MRI-guided biopsy on histologic examination.

The underestimation rate in DCIS was considered if a pathologically proven invasive carcinoma was seen at surgical excision or follow-up re-biopsy when the MRI-guided biopsy result was DCIS.

A biopsy was counted as technically successful if it was possible to see the target lesion on MRI on the day of the procedure, and the biopsy could be safely performed according to the performing physician.

Statistical Analysis

We performed descriptive statistics on our database using IBM SPSS Statistics for Windows, Version 28.0 (Armonk, NY). Qualitative variables were summarized by count and percentage, which included MRI indication, lesion type, and post-biopsy complications. Quantitative variables such as the average core number, age, follow-up time, and lesion size were reported as mean/median.

We tabulated numbers from all studies but some studies were excluded on a per-question basis when they did not report the numbers we were investigating. A random intercept logistic regression model was used to pool technical success rates, canceled biopsy rates, histopathology results, false-negative results, discordant rates, false-negative rates after excluding benign-discordant biopsies and upgrade rates in DCIS, ADH, and other high-risk lesion types. Weighted mean proportion and 95% confidence intervals (CIs) were reported. Of note, the random effects model uses weighted proportions, so: 1) pooled rates were not calculated by dividing the nominator by the denominator, 2) the denominators were different for each analysis, and 3) the pooled rates might not add up to 100%. Clopper-Pearson exact binomial intervals were calculated for each pooled proportion. Between-study heterogeneity was quantified by Higgin’s & Thompson’s I-squared statistic (25% low heterogeneity, 25–50% medium, >50% high) (17). Odds ratios were pooled using the random effects model.

Meta-regression with mixed-effects models was used to test the moderator effect of the year that the study was published (before or in 2010 versus after 2010), the average number of cores sampled (more than 13 cores sampled vs others), needle size (≤11G vs >11G) and mean lesion size (≤12 mm vs >12 mm) with the outcomes of false-negative rates, DCIS upgrade rates, ADH upgrades, and other high-risk lesions upgrade rates. The corresponding p-values were reported and p<0.05 was considered statistically significant. We used the R 4.2.1 (R core team, Vienna, Austria) and meta package (18).

Analyzed Data Cohort and Included Studies

A total of 318 abstracts were identified after the exclusion of the repeated articles. Of these 318 abstracts, 189 (59.43%) were excluded after title/abstract screening due to the title missing key research words (n = 177), not being an original article with novel data (n = 8), and using phantoms/models (n = 4). The remaining 129 studies (40.57%) were retrieved and 125 (39.31%) were reviewed in their full text. Seventy-five (25.58%) were excluded due to not being available fully in English (n = 15, 4.72%), missing technical factors of the MRI-guided biopsy (n = 6, 1.89%), not having histopathological outcomes (n = 38, 11.95%), being an interim result of an included study (n = 3, 0.94%) or lacking proper clinical/surgical/imaging follow-up (n = 13, 4.10%). The remaining 50 (15.72%) studies were included in this study and reviewed systematically (Figure 1). Table 1 summarizes the remaining 50 studies which met our inclusion criteria.

Figure 1
Table 1

The studies we included in this meta-analysis had an overall moderate to low risk of bias. Detailed information on the risk of biases of the studies included is shown in Figure 2.

Figure 2

Technical factors and biopsy success

Pooled reported data from 50 studies with 11,215 target lesions were reviewed. Varying magnet strength (1.5 or 3 Tesla), needle gauges (7–18), and needle types were used for biopsy.

Twenty-five studies out of 50 (50.00%) provided the number of recommended biopsies along with the number of successful ones. The rates were pooled using the random effects model. The pooled rate for canceled biopsies due to non-enhancement on the day of the procedure was 4.58% (95% CI: 1.81–11.11%) (Figure 3a). Canceled biopsies due to non-enhancement were excluded from the technical success analysis yielding a final technical success rate of 99.10% (95% CI: 97.89–99.62%) (Figure 3b).

A total of 11,087 successful MRI-guided biopsies were included in this review. A median of 13 cores (range: 2–60) was obtained per biopsy. Despite collecting enhancement kinetics data, these were not included in our analysis due to the insufficient number of studies describing lesion enhancement kinetics.

The number of patients was reported in 48/50 (96.0%) studies. In 10,463 successful biopsies in 7,893 women, the mean patient age was 51.8 years (range of mean/median, 45.5–58, standard deviation: ±2.8).

Indications for breast MRI

MRI indication information was available for 5,333 patients (5,333/7,893, 67.57%). The indication was breast cancer staging in 1,496 (28.05%, 95% CI: 26.85–29.28%), screening in 1,427 (26.76%, 95% CI: 25.57–27.97%), breast cancer surveillance in 1,027 (19.26%, 95% CI: 18.21–20.34%), diagnostic (abnormal mammogram/ultrasound or clinical symptoms) in 1,038 (19.46%, 95% CI: 18.41–20.55%), unknown primary in 74 (1.39%, 95% CI: 1.09–1.74%), and other in 271 (5.08%, 95% CI: 4.51–5.71%) (Supplementary Figure 1).

Figure 1

Those patients undergoing MRI for breast cancer surveillance [odds ratio (OR) 1.36 (95% CI: 0.96–1.93; p = 0.09)], diagnostic indication (OR 1.20, 95% CI: 0.87–1.67; p = 0.27) or breast cancer staging (OR 1.20, 95% CI: 0.79–1.82; p = 0.40) had higher rates of malignant outcomes. Of the MRI indications, fewer malignant outcomes were observed in screening (OR 0.47, 95% CI: 0.25–0.87; p = 0.02) (Figure 4).

Figure 4

Histopathology results and lesion types

Of 11,087 successful biopsies, the pooled rate for histopathology results was benign in 65.06% (95% CI: 59.15–70.54%), malignant in 29.64% (95% CI: 23.58–36.52%; invasive cancer, 15.16%, 95% CI: 12.56–18.18%; DCIS, 9.51%, 95% CI: 7.63–11.80%) and high risk in 16.69% (95% CI: 9.96–26.64%; ADH, 6.33%, 95% CI: 4.24–9.36%; other high-risk lesions, 12.73%, 95% CI: 7.12–21.73%) (Supplementary Figure 2). The pooled rate for invasive cancer among the malignant results was 62.10% (95% CI: 57.09–66.87%) and it was 40.00% (95% CI: 33.48–46.89%) for DCIS (Supplementary Figure 3). Among the high-risk lesions, the ADH pooled rate was 44.56% (95% CI: 30.84–59.15%) and the pooled rate for high-risk lesions other than ADH was 63.17% (95% CI: 51.40–73.55%) (Supplementary Figure 4).

Figure 2
Figure 4

Lesion type on MRI was available in 7,440 (67.11%) biopsies [5,145 mass (44.93%), 2,152 non-mass enhancement (18.79%), 143 focus (1.25%)]. The average mass enhancement size was 10.1 mm (range: 2–60) while the average non-mass enhancement size was 22.8 mm (range: 4–140), yielding an overall average lesion size of 12.4 mm (range: 2–140). Corresponding histopathological results were missing in 1,140 masses (1,140/5,145, 22.16%), 1,571 non-mass enhancements (1,571/2,152, 73.00%) and 82 foci (82/143, 57.34%). Of the 4,005 mass lesions, 2,417 (60.35%) were benign, 1,097 (27.39%) were malignant and 491 (12.26%) were high-risk. Overall, mass lesions were more likely to yield malignancy compared to non-mass and foci lesions [27.39% vs 11.36% (non-mass) and 18.03% foci, p<0.001]. Table 2 shows lesion types on MRI with the corresponding histopathology results.

Table 2

Upgrade Rates

Surgical upgrade to invasive cancer occurred in 12.24% of DCIS lesions (95% CI: 7.76–18.77%) (Figure 5a). The upgrade rate among all high-risk lesions was 15.14% (95% CI: 10.69–21.17%) (Figure 5b). Of 294 ADH lesions, upgrade to DCIS or invasive cancer was seen in 31.81% (95% CI: 25.57–38.77%) (Figure 5c) while a pooled upgrade rate of 6.75% (95% CI: 2.57–16.56%) (Figure 5d) was seen in high-risk lesions other than ADH (LCIS, ALH, FEA RSL/CSL). Among high-risk lesions, ADH had the highest upgrade rate to malignancy [OR 3.51 (95% CI: 2.18–5.65), p<0.001].

Figure 5

Benign-discordant biopsies and false negative rates

Short-term follow-up with a median of 25 months (range: 0.4–117) was performed in 1,651 (20.92%) patients. The pooled malignancy rate after the benign biopsy result was 1.64% (95% CI: 0.96–2.81%) (Figure 6a). The pooled radiology-pathology discordance rate was 2.48% (95% CI: 1.62–3.77%) (Figure 6b). When benign-discordant biopsies were excluded, the pooled false negative rate was 0.75% (95% CI: 0.34–1.62%) (Figure 6c).

When we compared studies based on the year published (before/in 2010 versus after 2010), needle size (≤11G vs >11G), the average number of cores sampled (more than 13 cores sampled vs others), and average lesion size (≤12 mm vs >12 mm), we did not find enough evidence to establish any association with the false negative rate, DCIS to invasive cancer or the high-risk lesion upgrade rate (p-values: 0.13–1.00). Table 3 represents the comparison results in detail.

Table 3

Post-biopsy Complications

Post-biopsy complications [158 (1.42%) hematoma, 17 (0.15%) vasovagal response, 19 (0.17%) other] were rare and seen in 186 out of 7,893 patients (2.36%, 95% CI: 2.03–2.72%).

References

1
Peters NH, Borel Rinkes IH, Zuithoff NP, Mali WP, Moons KG, Peeters PH. Meta-analysis of MR imaging in the diagnosis of breast lesions. Radiology 2008; 246: 116-124. (PMID: 18024435)
2
Spick C, Baltzer PA. Diagnostic utility of second-look US for breast lesions identified at MR imaging: systematic review and meta-analysis. Radiology 2014; 273: 401-409. (PMID: 25119022)
3
Plantade R, Thomassin-Naggara I. MRI vacuum-assisted breast biopsies. Diagn Interv Imaging 2014; 95: 779-801. (PMID: 24636589)
4
Mann RM, Balleyguier C, Baltzer PA, Bick U, Colin C, Cornford E, et al. Breast MRI: EUSOBI recommendations for women’s information. Eur Radiol 2015; 25: 3669-3678. (PMID: 26002130)
5
Sardanelli F, Boetes C, Borisch B, Decker T, Federico M, Gilbert FJ, et al. Magnetic resonance imaging of the breast: recommendations from the EUSOMA working group. Eur J Cancer 2010; 46: 1296-1316. (PMID: 20304629)
6
McGrath AL, Price ER, Eby PR, Rahbar H. MRI-guided breast interventions. J Magn Reson Imaging 2017; 46: 631-645. (PMID: 28470744)
7
Heywang-Köbrunner S, Heinig A, Pickuth D, Alberich T, Spielmann R. Interventional MRI of the breast: lesion localisation and biopsy. European radiology 2000; 10: 36-45. (PMID: 10663716)
8
Heywang-Köbrunner SH, Heinig A, Schaumlöffel U, Viehweg P, Buchmann J, Lampe D, et al. MR-guided percutaneous excisional and incisional biopsy of breast lesions. Eur Radiol 1999; 9: 1656-1665. (PMID: 10525886)
9
Viehweg P, Heinig A, Amaya B, Alberich T, Laniado M, Heywang-Köbrunner S. MR-guided interventional breast procedures considering vacuum biopsy in particular. Eur J Radiol 2002; 42: 32-39. (PMID: 12039018)
10
Perlet C, Heywang-Kobrunner SH, Heinig A, et al. Magnetic resonance-guided, vacuum-assisted breast biopsy: results from a European multicenter study of 538 lesions. Cancer 2006; 106: 982-990. (PMID: 16456807)
11
Liberman L, Bracero N, Morris E, Thornton C, Dershaw DD. MRI-guided 9-gauge vacuum-assisted breast biopsy: initial clinical experience. AJR Am J Roentgenol 2005; 185: 183-193. (PMID: 15972421)
12
Eby PR, Lehman CD. Magnetic resonance imaging-guided breast interventions. Top Magn Reson Imaging 2008; 19: 151-162. (PMID: 18941395)
13
Imschweiler T, Haueisen H, Kampmann G, Rageth L, Seifert B, Rageth C, et al. MRI-guided vacuum-assisted breast biopsy: comparison with stereotactically guided and ultrasound-guided techniques. Eur Radiol 2014; 24: 128-135. (PMID: 23979106)
14
Page MJ, McKenzie JE, Bossuyt PM, Boutron I, Hoffmann TC, Mulrow CD, et al. The PRISMA 2020 statement: an updated guideline for reporting systematic reviews. BMJ 2021; 372: n71. (PMID: 33782057)
15
Whiting PF, Rutjes AW, Westwood ME, Mallett S, Deeks JJ, Reitsma JB, et al. QUADAS-2: a revised tool for the quality assessment of diagnostic accuracy studies. Annals of internal medicine 2011; 155: 529-536. (PMID: 22007046)
16
The American Society of Breast Surgeons’ Consensus Guideline on Concordance Assessment of Image-Guided Breast Biopsies and Management of Borderline or High-Risk Lesions (2016) Accessed 7/1, 2022. https://www.breastsurgeons.org/docs/statements/Consensus-Guideline-on-Concordance-Assessment-of-Image-Guided-Breast-Biopsies.pdf?v2
17
Higgins JP, Thompson SG. Quantifying heterogeneity in a meta-analysis. Stat Med 2002; 21: 1539-1558. (PMID: 12111919)
18
Balduzzi S, Rücker G, Schwarzer G. How to perform a meta-analysis with R: a practical tutorial. Evid Based Ment Health 2019; 22: 153-160. (PMID: 31563865)
19
Pinkney DM, Chikarmane SA, Giess CS. Do benign-concordant breast MRI biopsy results require short interval follow-up imaging? Report of longitudinal study and review of the literature. Clin Imaging 2019; 57: 50-55. (PMID: 31129391)
20
Li J, Dershaw DD, Lee CH, Kaplan J, Morris EA. MRI follow-up after concordant, histologically benign diagnosis of breast lesions sampled by MRI-guided biopsy. AJR Am J Roentgenol 2009; 193: 850-855. (PMID: 19696301)
21
Huang ML, Speer M, Dogan BE, Rauch GM, Candelaria RP, Adrada BE, et al. Imaging-Concordant Benign MRI-Guided Vacuum-Assisted Breast Biopsy May Not Warrant MRI Follow-Up. AJR Am J Roentgenol 2017; 208: 916-922. (PMID: 28140609)
22
Lambert J, Steelandt T, Heywang-Köbrunner SH, Gieraerts K, Van Den Berghe I, Van Ongeval C, et al. Long-term MRI-guided vacuum-assisted breast biopsy results of 600 single-center procedures. Eur Radiol 2021; 31: 4886-4897. (PMID: 33459860)
23
Mahoney MC. Initial clinical experience with a new MRI vacuum-assisted breast biopsy device. J Magn Reson Imaging 2008; 28: 900-905. (PMID: 18821610)
24
Han BK, Schnall MD, Orel SG, Rosen M. Outcome of MRI-guided breast biopsy. AJR Am J Roentgenol 2008; 191: 1798-1804. (PMID: 19020252)
25
Malhaire C, El Khoury C, Thibault F, Athanasiou A, Petrow P, Ollivier L, et al. Vacuum-assisted biopsies under MR guidance: results of 72 procedures. Eur Radiol 2010; 20: 1554-1562. (PMID: 20119729)
26
Candelaria R, Fornage BD. Second‐look US examination of MR‐detected breast lesions. J Clin Ultrasound 2011; 39: 115-121. (PMID: 21387324)
27
Abe H, Schmidt RA, Shah RN, Shimauchi A, Kulkarni K, Sennett CA, et al. MR-directed (“Second-Look”) ultrasound examination for breast lesions detected initially on MRI: MR and sonographic findings. AJR Am J Roentgenol 2010; 194: 370-377. (PMID: 20093598)
28
Meissnitzer M, Dershaw DD, Lee CH, Morris EA. Targeted ultrasound of the breast in women with abnormal MRI findings for whom biopsy has been recommended. AJR Am J Roentgenol 2009; 193: 1025-1029. (PMID: 19770325)
29
Demartini WB, Eby PR, Peacock S, Lehman CD. Utility of targeted sonography for breast lesions that were suspicious on MRI. AJR Am J Roentgenol 2009; 192: 1128-1134. (PMID: 19304724)
30
Sim L, Hendriks J, Bult P, Fook-Chong S. US correlation for MRI-detected breast lesions in women with familial risk of breast cancer. Clinical radiology 2005; 60: 801-806. (PMID: 15978891)
31
Rauch GM, Dogan BE, Smith TB, Liu P, Yang WT. Outcome Analysis of 9-Gauge MRI-Guided Vacuum-Assisted Core Needle Breast Biopsies. AJR Am J Roentgenol 2012; 198: 292-299. (PMID: 22268171)
32
Winchester DJ, Bernstein JR, Jeske JM, Nicholson MH, Hahn EA, Goldschmidt RA, et al. Upstaging of atypical ductal hyperplasia after vacuum-assisted 11-gauge stereotactic core needle biopsy. Arch Surg 2003; 138: 619-623. (PMID: 12799332)
33
Jackman RJ, Birdwell RL, Ikeda DM. Atypical ductal hyperplasia: can some lesions be defined as probably benign after stereotactic 11-gauge vacuum-assisted biopsy, eliminating the recommendation for surgical excision? Radiology 2002; 224: 548-554. (PMID: 12147855)
34
Liberman L, Kaplan JB, Morris EA, Abramson AF, Menell JH, Dershaw DD. To Excise or To Sample the Mammographic Target: What Is the Goal of Stereotactic II-Gauge Vacuum-Assisted Breast Biopsy? AJR Am J Roentgeno 2002; 179: 679-683. (PMID: 12185043)
35
Liberman L, Cohen MA, Dershaw DD, Abramson AF, Hann LE, Rosen PP. Atypical ductal hyperplasia diagnosed at stereotaxic core biopsy of breast lesions: an indication for surgical biopsy. AJR Am J Roentgeno 1995; 164: 1111-1113. (PMID: 7717215)
36
Jackman RJ, Nowels KW, Shepard MJ, Finkelstein SI, Marzoni Jr FA. Stereotaxic large-core needle biopsy of 450 nonpalpable breast lesions with surgical correlation in lesions with cancer or atypical hyperplasia. Radiology 1994; 193: 91-95. (PMID: 8090927)
37
Liberman L. Percutaneous image-guided core breast biopsy. Radiologic Clinics 2002; 40: 483-500. (PMID: 12117188)
38
Heller SL, Elias K, Gupta A, Greenwood HI, Mercado CL, Moy L. Outcome of high-risk lesions at MRI-guided 9-gauge vacuum- assisted breast biopsy. AJR Am J Roentgenol 2014; 202: 237-245. (PMID: 24370150)
39
Lourenco AP, Khalil H, Sanford M, Donegan L. High-risk lesions at MRI-guided breast biopsy: frequency and rate of underestimation. AJR Am J Roentgenol 2014; 203: 682-686. (PMID: 25148176)
40
Weinfurtner RJ, Patel B, Laronga C, Lee MC, Falcon SL, Mooney BP, et al. Magnetic resonance imaging-guided core needle breast biopsies resulting in high-risk histopathologic findings: upstage frequency and lesion characteristics. Clin Breast Cancer 2015; 15: 234-239. (PMID: 25579460)
41
Verheyden C, Pages-Bouic E, Balleyguier C, Cherel P, Lepori D, Laffargue G, et al. Underestimation Rate at MR Imaging-guided Vacuum-assisted Breast Biopsy: A Multi-Institutional Retrospective Study of 1509 Breast Biopsies. Radiology 2016; 281: 708-719. (PMID: 27355898)
42
Speer ME, Huang ML, Dogan BE, Adrada BE, Candelaria RP, Hess KR, et al. High risk breast lesions identified on MRI-guided vacuum-assisted needle biopsy: outcome of surgical excision and imaging follow-up. Br J Radiol 2018; 91: 20180300. (PMID: 29947265)
43
Rageth CJ, O’Flynn EAM, Pinker K, Kubik-Huch RA, Mundinger A, Decker T, et al. Second International Consensus Conference on lesions of uncertain malignant potential in the breast (B3 lesions). Breast Cancer Res Treat 2019; 174: 279-296. (PMID: 30506111)
44
Pinder SE, Shaaban A, Deb R, Desai A, Gandhi A, Lee AHS, et al. NHS Breast Screening multidisciplinary working group guidelines for the diagnosis and management of breast lesions of uncertain malignant potential on core biopsy (B3 lesions). Clin Radiol 2018; 73: 682-692. (PMID: 29773220)
45
Lehman CD, Deperi ER, Peacock S, McDonough MD, Demartini WB, Shook J. Clinical experience with MRI-guided vacuum-assisted breast biopsy. AJR Am J Roentgenol 2005; 184: 1782-1787. (PMID: 15908530)
46
Liberman L, Morris EA, Dershaw DD, Thornton CM, Van Zee KJ, Tan LK. Fast MRI-guided vacuum-assisted breast biopsy: initial experience. AJR Am J Roentgenol 2003; 181: 1283-1293. (PMID: 14573421)
47
Orel SG, Rosen M, Mies C, Schnall MD. MR imaging-guided 9-gauge vacuum-assisted core-needle breast biopsy: initial experience. Radiology 2006; 238: 54-61. (PMID: 16304093)
48
Gebauer B, Bostanjoglo M, Moesta KT, Schneider W, Schlag PM, Felix R. Magnetic resonance-guided biopsy of suspicious breast lesions with a handheld vacuum biopsy device. Acta Radiol 2006; 47: 907-913. (PMID: 17077039)
49
Ambinder EB, Myers KS, Mullen LA, Di Carlo P, Philip M, Fragomeni R, et al. Breast MRI biopsy cancellation due to lesion nonvisualization. Breast J 2020; 26: 2021-2025. (PMID: 32924192)
50
Pinnamaneni N, Moy L, Gao Y, Melsaether AN, Babb JS, Toth HK, et al. Canceled MRI-guided Breast Biopsies Due to Nonvisualization: Follow-up and Outcomes. Acad Radiol 2018; 25: 1101-1110. (PMID: 29478921)
51
Hefler L, Casselman J, Amaya B, Heinig A, Alberich T, Koelbl H, et al. Follow-up of breast lesions detected by MRI not biopsied due to absent enhancement of contrast medium. Eur Radiol 2003; 13: 344-346. (PMID: 12599000)
52
Viehweg P, Bernerth T, Kiechle M, Buchmann J, Heinig A, Koelbl H, et al. MR-guided intervention in women with a family history of breast cancer. Eur J Radiol 2006; 57: 81-89. (PMID: 16364583)
53
Brennan SB, Sung JS, Dershaw DD, Liberman L, Morris EA. Cancellation of MR imaging-guided breast biopsy due to lesion nonvisualization: frequency and follow-up. Radiology 2011; 261: 92-99. (PMID: 21852565)
54
Johnson KS, Baker JA, Lee SS, Soo MS. Cancelation of MRI guided breast biopsies for suspicious breast lesions identified at 3.0 T MRI: reasons, rates, and outcomes. Acad Radiol 2013; 20: 569-575. (PMID: 23473719)
55
Niell BL, Lee JM, Johansen C, Halpern EF, Rafferty EA. Patient outcomes in canceled MRI-guided breast biopsies. AJR Am J Roentgenol 2014; 202: 223-228. (PMID: 24370148)
56
Lee JM, Kaplan JB, Murray MP, Bartella L, Morris EA, Joo S, et al. Imaging–histologic discordance at MRI-guided 9-gauge vacuum-assisted breast biopsy. AJR Am J Roentgenol 2007; 189: 852-859. (PMID: 17885056)
57
Schueller G, Schueller-Weidekamm C, Helbich T. Accuracy of ultrasound-guided, large-core needle breast biopsy. Eur Radiol 2008; 18: 1761-1773. (PMID: 18414872)
58
Venkataraman S, Dialani V, Gilmore HL, Mehta TS. Stereotactic core biopsy: comparison of 11 gauge with 8 gauge vacuum assisted breast biopsy. Eur J Radiol 2012; 81: 2613-2619. (PMID: 22127375)
59
Liberman L, Drotman M, Morris EA, LaTrenta LR, Abramson AF, Zakowski MF, et al. Imaging–histologic discordance at percutaneous breast biopsy: an indicator of missed cancer. Cancer 2000; 89: 2538-2546. (PMID: 11135213)
60
Jackman RJ, Marzoni Jr FA, Rosenberg J. False-negative diagnoses at stereotactic vacuum-assisted needle breast biopsy: long-term follow-up of 1,280 lesions and review of the literature. AJR Am J Roentgenol 2009; 192: 341-351. (PMID: 19155393)
61
Lewin AA, Heller SL, Jaglan S, Elias K, Newburg A, Melsaether A, et al. Radiologic-Pathologic Discordance and Outcome After MRI-Guided Vacuum-Assisted Biopsy. AJR Am J Roentgenol 2016; 208: W17-W22. (PMID: 27786559)
62
Zhang C, Lewis DR, Nasute P, Hayes M, Warren LJ, Gordon PB. The negative predictive value of ultrasound-guided 14-gauge core needle biopsy of breast masses: a validation study of 339 cases. Cancer Imaging 2012; 12: 488-496. (PMID: 23113970)
63
Youk JH, Kim EK, Kim MJ, Ko KH, Kwak JY, Son EJ, et al. Concordant or discordant? Imaging-pathology correlation in a sonography-guided core needle biopsy of a breast lesion. Korean J Radiol 2011; 12: 232-240. (PMID: 21430941)
64
Schrading S, Strobel K, Keulers A, Dirrichs T, Kuhl CK. Safety and Efficacy of Magnetic Resonance-Guided Vacuum-Assisted Large-Volume Breast Biopsy (MR-Guided VALB). Invest Radiol 2017; 52: 186-193. (PMID: 27861232)
65
Spick C, Schernthaner M, Pinker K, Kapetas P, Bernathova M, Polanec SH, et al. MR-guided vacuum-assisted breast biopsy of MRI-only lesions: a single center experience. Eur Radiol 2016; 26: 3908-3916. (PMID: 26984430)
66
Belloni E, Panizza P, Ravelli S, De Cobelli F, Gusmini S, Losio C, et al. MR-guided stereotactic breast biopsy using a mixed ferromagnetic-nonmagnetic coaxial system with 12- to 18-gauge needles: clinical experience and long-term outcome. Radiol Med 2013; 118: 1137-1148. (PMID: 23801401)
67
Shaylor SD, Heller SL, Melsaether AN, Gupta D, Gupta A, Babb J, et al. Short interval follow-up after a benign concordant MR-guided vacuum assisted breast biopsy--is it worthwhile? Eur Radiol 2014; 24: 1176-1185. (PMID: 24623366)
68
Dratwa C, Jalaguier-Coudray A, Thomassin-Piana J, Gonin J, Chopier J, Antoine M, et al. Breast MR biopsy: Pathological and radiological correlation. Eur Radiol 2016; 26: 2510-2519. (PMID: 26511630)
69
Friedman P, Enis S, Pinyard J. Magnetic resonance imaging-guided vacuum-assisted breast biopsy: an initial experience in a community hospital. Can Assoc Radiol J 2009; 60: 196-200. (PMID: 19782886)
70
Hayward JH, Ray KM, Wisner DJ, Joe BN. Follow-up outcomes after benign concordant MRI-guided breast biopsy. Clin Imaging 2016; 40: 1034-1039. (PMID: 27348060)
71
Lee JM, Kaplan JB, Murray MP, Liberman L. Complete excision of the MRI target lesion at MRI-guided vacuum-assisted biopsy of breast cancer. AJR Am J Roentgenol 2008; 191: 1198-1202. (PMID: 18806165)
72
An Y, Kim S, Kang B, Lee J. Usefulness of magnetic resonance imaging-guided vacuum-assisted breast biopsy in Korean women: a pilot study. World J Surg Oncol 2013; 11: 200. (PMID: 23948057)
73
Bahrs SD, Hattermann V, Preibsch H, Hahn M, Staebler A, Claussen CD, et al. MR imaging-guided vacuum-assisted breast biopsy: reduction of false-negative biopsies by short-term control MRI 24-48 h after biopsy. Clin Radiol 2014; 69: 695-702. (PMID: 24679372)
74
Carbognin G, Girardi V, Brandalise A, Baglio I, Bucci A, Bonetti F, et al. MR-guided vacuum-assisted breast biopsy in the management of incidental enhancing lesions detected by breast MR imaging. Radiol Med 2011; 116: 876-885. (PMID: 21293942)
75
Chen X, Lehman CD, Dee KE. MRI-guided breast biopsy: clinical experience with 14-gauge stainless steel core biopsy needle. AJR Am J Roentgenol 2004; 182: 1075-1080. (PMID: 15039191)
76
Crystal P, Sadaf A, Bukhanov K, McCready D, O’Malley F, Helbich TH. High-risk lesions diagnosed at MRI-guided vacuum-assisted breast biopsy: can underestimation be predicted? Eur Radiol 2011; 21: 582-589. (PMID: 20839000)
77
Dogan BE, Le-Petross CH, Stafford JR, Atkinson N, Whitman GJ. MRI-guided vacuum-assisted breast biopsy performed at 3 T with a 9-gauge needle: preliminary experience. AJR Am J Roentgenol 2012; 199: W651-653. (PMID: 23096211)
78
Ferré R, Ianculescu V, Ciolovan L, Mathieu MC, Uzan C, Canale S, et al. Diagnostic Performance of MR-guided Vacuum-Assisted Breast Biopsy: 8 Years of Experience. Breast J 2016; 22: 83-89. (PMID: 26511082)
79
Ghate SV, Rosen EL, Soo MS, Baker JA. MRI-guided vacuum-assisted breast biopsy with a handheld portable biopsy system. AJR Am J Roentgenol 2006; 186: 1733-1736. (PMID: 16714667)
80
Hauth EA, Jaeger HJ, Lubnau J, Maderwald S, Otterbach F, Kimmig R, et al. MR-guided vacuum-assisted breast biopsy with a handheld biopsy system: clinical experience and results in postinterventional MR mammography after 24 h. Eur Radiol 2008; 18: 168-176. (PMID: 17609959)
81
Jung HN, Han BK, Ko EY, Shin JH. Initial experience with magnetic resonance-guided vacuum-assisted biopsy in korean women with breast cancer. J Breast Cancer 2014; 17: 270-278. (PMID: 25320626)
82
Kılıç F, Eren A, Tunç N, et al. Magnetic Resonance Imaging Guided Vacuum Assisted and Core Needle Biopsies. J Breast Health 2016; 12: 25-30. (PMID: 28331727)
83
Lee JM, Kaplan JB, Murray MP, et al. Underestimation of DCIS at MRI-guided vacuum-assisted breast biopsy. AJR Am J Roentgenol 2007; 189: 468-474. (PMID: 17646475)
84
Lee SJ, Mahoney MC, Redus Z. The Management of Benign Concordant MRI-guided Brest Biopsies: Lessons Learned. Breast J 2015; 21: 665-668. (PMID: 26385013)
85
Lehman CD, Eby PR, Chen X, Dee KE, Thursten B, McCloskey J. MR imaging-guided breast biopsy using a coaxial technique with a 14-gauge stainless steel core biopsy needle and a titanium sheath. AJR Am J Roentgenol 2003; 181: 183-185. (PMID: 12818854)
86
Liberman L, Holland AE, Marjan D, et al. Underestimation of atypical ductal hyperplasia at MRI-guided 9-gauge vacuum-assisted breast biopsy. AJR Am J Roentgenol 2007; 188: 684-690. (PMID: 17312054)
87
Meeuwis C, Veltman J, van Hall HN, et al. MR-guided breast biopsy at 3T: diagnostic yield of large core needle biopsy compared with vacuum-assisted biopsy. European radiology 2012; 22: 341-349. (PMID: 21915606)
88
Myers KS, Kamel IR, Macura KJ. MRI-guided breast biopsy: outcomes and effect on patient management. Clin Breast Cancer 2015; 15: 143-152. (PMID: 25499596)
89
Noroozian M, Gombos EC, Chikarmane S, et al. Factors that impact the duration of MRI-guided core needle biopsy. AJR Am J Roentgenol 2010; 194: W150-157. (PMID: 20093566)
90
O’Connor V, Arena E, Albright J, et al. Histological assessment of breast lesions identified exclusively by magnetic resonance. Am Surg 2014; 80: 944-947. (PMID: 25264635)
91
Perretta T, Pistolese CA, Bolacchi F, Cossu E, Fiaschetti V, Simonetti G. MR imaging-guided 10-gauge vacuum-assisted breast biopsy: histological characterisation. Radiol Med 2008; 113: 830-840. (PMID: 18633687)
92
Peters W, Fornasier V. Complications from injectable materials used for breast augmentation. Can J Plast Surg 2009; 17: 89-96. (PMID: 20808751)
93
Schrading S, Simon B, Braun M, Wardelmann E, Schild HH, Kuhl CK. MRI-guided breast biopsy: influence of choice of vacuum biopsy system on the mode of biopsy of MRI-only suspicious breast lesions. AJR Am J Roentgenol 2010; 194: 1650-1657. (PMID: 20489109)
94
Tozaki M, Yamashiro N, Sakamoto M, Sakamoto N, Mizuuchi N, Fukuma E. Magnetic resonance-guided vacuum-assisted breast biopsy: results in 100 Japanese women. Jpn J Radiol 2010; 28: 527-533. (PMID: 20799018)
95
Zebic-Sinkovec M, Hertl K, Kadivec M, Cavlek M, Podobnik G, Snoj M. Outcome of MRI-guided vacuum-assisted breast biopsy - initial experience at Institute of Oncology Ljubljana, Slovenia. Radiol Oncol 2012; 46: 97-105. (PMID: 23077445)